Lumen-apposing Metal Stent for the drainage of pancreatic pseudocyst or a gallbladder through a transgastric or transduodenal approach. Pre-loaded on the Electrocautery Delivery System.
• Prevents migration and maintains anastomotic conduit between two adjacent organs
• The flexible flare design helps accommodative apposition regardless of the wall thickness
• 8, 10, 16mm diameters enable to apply to various indications
• Select the pad specified by generator manufacturer
• Set to Pure Cut Mode 80-120 Watts, 400-500Vp
• Connect the plug to the electrosurgical unit with a compatible active cable
• Step 1. When the outer X-ray marker overlaps with the inner X-ray marker
→ The distal flange is completely opened
• Step 2. When the blue marker is visible under endoscopic view
→ Start opening proximal flange